WU XUEZHOU;WANG ZAIYONG National Institute of Metrology, Beijing. DETERMINATION OF DEAD TIME AND COINCIDENCE RESOLVING TIME WITH DOUBLE OSCILLATOR[J]. Atomic Energy Science and Technology, 1986, 20(2): 235-235. DOI: 10.7538/yzk.1986.20.02.0235
Citation: WU XUEZHOU;WANG ZAIYONG National Institute of Metrology, Beijing. DETERMINATION OF DEAD TIME AND COINCIDENCE RESOLVING TIME WITH DOUBLE OSCILLATOR[J]. Atomic Energy Science and Technology, 1986, 20(2): 235-235. DOI: 10.7538/yzk.1986.20.02.0235

DETERMINATION OF DEAD TIME AND COINCIDENCE RESOLVING TIME WITH DOUBLE OSCILLATOR

  • At present, the double oscillator method is among the best methods formeasuring dead time and coincidence resolving time. This paper introduces themeasurement principle of this method. The standard deviation of measuring deadtime and resolving time, with this method, is respectly about 1×10~(-4) and(2--3)×10~(-4).