YIN Zai-zhe1,P.R.Hugtenburg2, A.H.Beddoe2(1.Institute of Nuclear and New Energy Technology, Tsinghua University, Beijing 100084, China;2.Queen Elizabeth Hosiptal, Birmingham B15 2TH, UK). Effect of Low Energy Scatter Photon on Solid State Detectorsin Measurement of Water Absorbed Dose[J]. Atomic Energy Science and Technology, 2004, 38(S1): 236-236. DOI: 10.7538/yzk.2004.38.S1.0236
Citation: YIN Zai-zhe1,P.R.Hugtenburg2, A.H.Beddoe2(1.Institute of Nuclear and New Energy Technology, Tsinghua University, Beijing 100084, China;2.Queen Elizabeth Hosiptal, Birmingham B15 2TH, UK). Effect of Low Energy Scatter Photon on Solid State Detectorsin Measurement of Water Absorbed Dose[J]. Atomic Energy Science and Technology, 2004, 38(S1): 236-236. DOI: 10.7538/yzk.2004.38.S1.0236
  • The measurements of scatter factor are carried out using diode and RK ionization chamber for 6 and 15 MV medical linear accelerator energy photons in water phantom. The photon energy spectra for 6 and 15 MV in water phantom are also calculated using Monte-Carlo method. The aim of this paper is to correct the error resulting in the overdosing of diode detector due to low energy scatter photons. The paper also supports the quality assurance(QA)and quality control(QC) in radiotherapy physics.