Analysis of Gas-Filled Time-of-Flight Detection Method[J]. Atomic Energy Science and Technology, 2006, 40(2): 145-149. DOI: 10.7538/yzk.2006.40.02.0145
Citation: Analysis of Gas-Filled Time-of-Flight Detection Method[J]. Atomic Energy Science and Technology, 2006, 40(2): 145-149. DOI: 10.7538/yzk.2006.40.02.0145

Analysis of Gas-Filled Time-of-Flight Detection Method

More Information
  • Received Date: May 31, 2005
  • Revised Date: October 13, 2005
  • The paper concerns in the principle and the experimental results of gas-filled timeofflight (GF-TOF) detector. 36Cl and 36S particles were successfully separated in the TOF spectra from the GF-TOF detector for the ion with the energy of 64, 48 and 33 MeV, respectively. The comparison of the separation power of isobars between the GF-TOF method and the ΔE-E method was carried out. The experimental results indicate that the separation power of the ΔE-E method is better than that of the GF-TOF method at the energy larger than 40 MeV, and the GF-TOF method is better than the ΔE-E method at the energy less than 40 MeV, especially at the energy range of 20-40 MeV.
  • [1]
    JIANG Shan, HE Ming, WU Shaoyong, et al. A gas-filled time-of-flight detector system and a new particle identification[J]. Nucl Instrum Methods, 2000,B172:305-309.
    [2]
    姜山,何明,蒋崧生, 等. 一台用于加速器质谱测量的充气飞行时间探测器[J]. 核技术,2000,23(10):718-720.
    JIANG Shan, HE Ming, JIANG Songsheng, et al. A gasfilled timeofflight detector for accelerator mass spectrometric measurements[J]. Nuclear Techniques, 2000:23(10):718-720(in Chinese).
    [3]
    RUAN Xiangdong, GUAN Yongjing, HE Ming, et al. A method for isobaric identification with gasfilled TOF detector system[J]. Nucl Instrum Methods, 2004, B223-224: 172-175.
    [4]
    GUAN Yongjing, RUAN Xiangdong, HE Ming, et al. Isobaric identification using gas-filled time-of-flight measurements in an accelerator mass spectrometry[J]. Chin Phys Lett, 2005: 22(7):1 622-1 624.