WEI Xing-jian, DENG Da-chao, YU Chun-rong, YUAN Li. Study on Calibration Technique of System Deviation of Multi-collector Thermal Ionization Mass Spectrometry[J]. Atomic Energy Science and Technology, 2007, 41(增刊2): 413-417. DOI: 10.7538/yzk.2007.41.suppl2.0413
Citation: WEI Xing-jian, DENG Da-chao, YU Chun-rong, YUAN Li. Study on Calibration Technique of System Deviation of Multi-collector Thermal Ionization Mass Spectrometry[J]. Atomic Energy Science and Technology, 2007, 41(增刊2): 413-417. DOI: 10.7538/yzk.2007.41.suppl2.0413

Study on Calibration Technique of System Deviation of Multi-collector Thermal Ionization Mass Spectrometry

More Information
  • Received Date: November 14, 2007
  • Revised Date: November 25, 2007
  • The system deviation of multi-collector thermal ionization mass spectrometry was researched, and the calibration techniques were established. The mass fractionation, multi-collector efficiency and the amplifier gain were calibrated by using isotope reference material and total evaporation technique, peak jumping technique, standard voltage and virtual amplifier technique, espectively. The new mathematical model for calibrating the nonlinearity of the amplifiers and ion count system was also established, and the calibration techniques were put forward. Based on these techniques, the isotope ratios can be measured accurately by multi-collector mass spectrometry.
  • [1]
    De OLIVERRA O P, De BOLLE W, RICHTER S, et al. Evaluation of the n (235U)/ n (238U) isotope ratio measurements in a set of uranium samples by thermal ionization mass spectrometry[J]. International Journal of Mass Spectrometry, 2005, 246: 35-42.
    [2]
    RICHTER S, GOLDBERG S A. Improved techniques for high accuracy isotope radio measurements of nuclear materials using thermal ionization mass spectrometry[J]. International Journal of Mass Spectrometry, 2003, 229: 181-197.
    [3]
    CHU Zhuyin, YANG Yueheng, QIAO Guangsheng. A calculation method to eliminate gain effect on isotopic measurement using the virtual amplifier multicollector mass spectrometer[J]. International Journal of Mass Spectrometry, 2006, 253: 130-135.
    [4]
    RICHTER S, GOLDBERG S A, MASON P B, et al. Linearity test for secondary multipliers used in isotope ratio mass spectrometry[J]. International Journal of Mass Spectrometry, 2001, 206: 105-127.

Catalog

    Article views (1545) PDF downloads (1230) Cited by()

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return