QI Ming-sen, ZHOU Chang-gui, LIU Xian-zhong, WANG Xu. Measurement of Specific Activity of 169Yb Radioactive Solution[J]. Atomic Energy Science and Technology, 2008, 42(3): 227-229. DOI: 10.7538/yzk.2008.42.03.0227
Citation: QI Ming-sen, ZHOU Chang-gui, LIU Xian-zhong, WANG Xu. Measurement of Specific Activity of 169Yb Radioactive Solution[J]. Atomic Energy Science and Technology, 2008, 42(3): 227-229. DOI: 10.7538/yzk.2008.42.03.0227

Measurement of Specific Activity of 169Yb Radioactive Solution

  • The paper brifely describes the principle and measurement technology on measuring specific activity of 169Yb radioactive solution by 4πX(PPC)-γ and 4πX(L•S)-γ activity standard apparatus. The method using extrapolation to select coincidence resolving time was given out. The measurement results from different standard apparatus were compared. The deviation is within 0.5%.