ZHU Li-qun, MENG Yan-tai, LIU Da-ming, CAI Ding-kan, WANG Xiao-zhong, BAI Lei, XU Xiao-ming, JIA Xiang-jun, SHEN Ning, GAN Lin. Mass Measurement of Plutonium With Calorimetric Method[J]. Atomic Energy Science and Technology, 2008, 42(6): 497-500. DOI: 10.7538/yzk.2008.42.06.0497
Citation: ZHU Li-qun, MENG Yan-tai, LIU Da-ming, CAI Ding-kan, WANG Xiao-zhong, BAI Lei, XU Xiao-ming, JIA Xiang-jun, SHEN Ning, GAN Lin. Mass Measurement of Plutonium With Calorimetric Method[J]. Atomic Energy Science and Technology, 2008, 42(6): 497-500. DOI: 10.7538/yzk.2008.42.06.0497

Mass Measurement of Plutonium With Calorimetric Method

  • In nuclear safeguards field, calorimeter is one of the most accurate and convenient nondestructive instruments for nuclear material accountancy. A calorimeter designed to measure plutonium in storage containers with thermal powers ranging from 0.1 to 15 W was used to measure the mass of 238Pu in three samples. Measured results show that the relative deviation between the measured 238Pu mass and standard reference 238Pu mass in samples of 238Pu is -1.7%.