WANG Qun-shu, KANG Ke-jun, SONG Zhao-hui, LI Gang, XIA Liang-bin. Experimental Study on Energy Sensitivity of ST401 Scintillation Detector to Gamma Ray[J]. Atomic Energy Science and Technology, 2008, 42(9): 856-860. DOI: 10.7538/yzk.2008.42.09.0856
Citation: WANG Qun-shu, KANG Ke-jun, SONG Zhao-hui, LI Gang, XIA Liang-bin. Experimental Study on Energy Sensitivity of ST401 Scintillation Detector to Gamma Ray[J]. Atomic Energy Science and Technology, 2008, 42(9): 856-860. DOI: 10.7538/yzk.2008.42.09.0856

Experimental Study on Energy Sensitivity of ST401 Scintillation Detector to Gamma Ray

  • The energy sensitivities of a current detector made up of ST401 plastic scintillator widely used in the field of pulse gamma rays measurements were studied experimentally. By means of pulse reactor and accelerator, a series of gamma rays with single or quasi-single energy were produced for calibration. The detector was calibrated by thirteen gamma energies including 137Cs and 60Co. The curve of relative energy sensitivities of the detector to incident gamma rays was obtained. The experimental result was compared with the theoretical calculation and they are in agreement within uncertainty.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return