PAN Hong-bo, OUYANG Xiao-ping, ZHANG Xian-peng, LI Da-hai, LI Hong-yun, QIU Meng-tong, LI Gang. DirectRadiation Response of CdZnTe Detector to 14 MeV Pulsed Neutron Beam[J]. Atomic Energy Science and Technology, 2009, 43(4): 1-315. DOI: 10.7538/yzk.2009.43.04.0381
Citation: PAN Hong-bo, OUYANG Xiao-ping, ZHANG Xian-peng, LI Da-hai, LI Hong-yun, QIU Meng-tong, LI Gang. DirectRadiation Response of CdZnTe Detector to 14 MeV Pulsed Neutron Beam[J]. Atomic Energy Science and Technology, 2009, 43(4): 1-315. DOI: 10.7538/yzk.2009.43.04.0381

DirectRadiation Response of CdZnTe Detector to 14 MeV Pulsed Neutron Beam

  • CdZnTe (CZT) crystal is one of the most attractive materials used for the new type radiation detector due to its excellent material properties. The directradiation response of CZT detector to 14 MeV neutron was calculated by MonteCarlo code MCNP. The measurement of the CZT detector to 14 MeV neutron was performed on ING103 pulsed neutron generator. The results indicate that the theoretical value accords with the experimental result. When coverage factor is 2, the expand uncertainty is less than 14%. With the sensitivity of 1.28×10-16 C·cm2 to 14 MeV neutron, the CZT detector can be applied to the highintensity pulsed neutron measurement.