YAO Xiang-wu, YANG Yan-yun, WANG Jian-song, HUANG Mei-rong, GAO Qi, GUO Zhong-yan, XIAO Guo-qing, XU Hu-shan, HU Zheng-guo, YU Yu-hong, QIU Hao, YAN Xin-shuai. Development of 8×8 CsI(Tl) Array Detector[J]. Atomic Energy Science and Technology, 2010, 44(3): 358-361. DOI: 10.7538/yzk.2010.44.03.0358
Citation: YAO Xiang-wu, YANG Yan-yun, WANG Jian-song, HUANG Mei-rong, GAO Qi, GUO Zhong-yan, XIAO Guo-qing, XU Hu-shan, HU Zheng-guo, YU Yu-hong, QIU Hao, YAN Xin-shuai. Development of 8×8 CsI(Tl) Array Detector[J]. Atomic Energy Science and Technology, 2010, 44(3): 358-361. DOI: 10.7538/yzk.2010.44.03.0358
  • The principle and structure of a 8×8 CsI(Tl) array detector were described in this paper. For each unit, there are a CsI(Tl) crystal, an optical conductor and a photoelectric multiplier tube. The length of CsI(Tl) crystal is about 50 mm with a 21 mm×21 mm front surface and a 23.1 mm×23.1 mm back surface. The energy resolution is 2.7% for 30 MeV proton and 1.5% for 170 MeV 7Be, which was tested on the RIBLL (Radioactive Ion Beam Line in Lanzhou). It can be used to identify the charged particles in the nuclear experiments induced by the radioactive beam.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return