ZHONG Zhen-yuan, LI Mou, DAI Yi-hua, ZHANG Hai-tao, LIU Jie, CHANG Yong-fu. A Method for Specific Activity Measurement of 241Am Solution[J]. Atomic Energy Science and Technology, 2010, 44(5): 523-526. DOI: 10.7538/yzk.2010.44.05.0523
Citation: ZHONG Zhen-yuan, LI Mou, DAI Yi-hua, ZHANG Hai-tao, LIU Jie, CHANG Yong-fu. A Method for Specific Activity Measurement of 241Am Solution[J]. Atomic Energy Science and Technology, 2010, 44(5): 523-526. DOI: 10.7538/yzk.2010.44.05.0523

A Method for Specific Activity Measurement of 241Am Solution

  • According to the principle of coincidence theory, the specific activity of 241Am solution was determined on 4πα+4πγ counting standard device by γ efficiency extrapolation, and the problems of constant correction coefficients of self-absorption and scattering in α ionization chamber method were solved. The method was based on the alteration of detection efficiency when the height of elevator was altered, and the activity was obtained by γ fitting extrapolation according to detection efficiency. The results of more than 20 alpha radioactive sources by this method in our work are accordant with those of 2πα ionization chamber, and their uncertainties are improved to 0.4%.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return