Calculation of Geometric Factor on Condition for α Detector and Surface Source Being Not Coaxial
-
Graphical Abstract
-
Abstract
A geometric factor was calculated using projection method on condition for α detector and surface source being not coaxial, and compared with experimental result. The differences between the geometric factor and experimental result are mainly consistent. The study is beneficial to α spectrometer calibration and uncertainty evaluation result from the geometric position in the routine measurement.
-
-