CHEN Qiu-qiu, CHEN Yong, YU Hong-liang, SUN Liang. Calculation of Scatter Calibration Factor for 60Co Therapy Unit With Monte-Carlo Method[J]. Atomic Energy Science and Technology, 2011, 45(2): 230-234. DOI: 10.7538/yzk.2011.45.02.0230
Citation: CHEN Qiu-qiu, CHEN Yong, YU Hong-liang, SUN Liang. Calculation of Scatter Calibration Factor for 60Co Therapy Unit With Monte-Carlo Method[J]. Atomic Energy Science and Technology, 2011, 45(2): 230-234. DOI: 10.7538/yzk.2011.45.02.0230

Calculation of Scatter Calibration Factor for 60Co Therapy Unit With Monte-Carlo Method

  • Three scatter calibration factors of 60Co therapy unit were calculated by using Monte-Carlo method. The effects of radiation field and depth to total scatter calibration factor Sc,p and the phantom scatter calibration factor Sp of 60Co therapy unit were also calculated. The calculating results agree well with the measuring results. Sc,p and Sp increase with the increasing of the fields. Sc,p and Sp have increasing trend with depth increasing while beam field larger than 10 cm×10 cm. Sc,p and Sp have decreasing trend with depth increasing while beam field smaller than 10 cm×10 cm. Scatter calibration factor is not be overlooked in calculating the prescription dose. Monte-Carlo method is able to establish a set of accurate and comprehensive scatter calibration factor, and provides the basis of clinical use, quality assurance and quality control for radiotherapy.