CHEN Guo-yun, WEI Zhi-yong, XIN Yong, FANG Mei-hua, HUANG San-bo, HUANG Guo-qing. Detection Efficiency and Sensitivity of Boron-Lined Ionization Chamber for Neutron[J]. Atomic Energy Science and Technology, 2011, 45(2): 244-249. DOI: 10.7538/yzk.2011.45.02.0244
Citation: CHEN Guo-yun, WEI Zhi-yong, XIN Yong, FANG Mei-hua, HUANG San-bo, HUANG Guo-qing. Detection Efficiency and Sensitivity of Boron-Lined Ionization Chamber for Neutron[J]. Atomic Energy Science and Technology, 2011, 45(2): 244-249. DOI: 10.7538/yzk.2011.45.02.0244

Detection Efficiency and Sensitivity of Boron-Lined Ionization Chamber for Neutron

  • Formulas for detection efficiency and sensitivity of plate boron-lined ionization chamber were derived from its working principle, and their values for thermal neutrons were obtained. The saturated value of the detection efficiency for thermal neutrons of ionization chamber is 1.35% and the sensitivity is 9.65×10-14 A•cm-2•s-1, which approximately coincides with the result of 8.43×10-14 A•cm-2•s-1 given by a known formula. Contribution of α particles to the detection efficiency shows small difference with Li ions while that to the sensitivity is predominating. A suitable thickness of boron film, slowing fast neutron and using enriched boron can improve both detection efficiency and sensitivity of boron-lined ionization chamber.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return