LI Xin-nian, FAN Xiang-xiang, HUANG Min, FANG Xiao-ming, QIAN Meng-lu, CHENG Qian. Neutron Detection System for Nuclear Effect of Acoustic Cavitation[J]. Atomic Energy Science and Technology, 2011, 45(5): 606-610. DOI: 10.7538/yzk.2011.45.05.0606
Citation: LI Xin-nian, FAN Xiang-xiang, HUANG Min, FANG Xiao-ming, QIAN Meng-lu, CHENG Qian. Neutron Detection System for Nuclear Effect of Acoustic Cavitation[J]. Atomic Energy Science and Technology, 2011, 45(5): 606-610. DOI: 10.7538/yzk.2011.45.05.0606

Neutron Detection System for Nuclear Effect of Acoustic Cavitation

  • The neutron detection system based on ST-451 liquid scintillator detector and BF3 proportional counters was arranged, the instrument precision of the neutron spectrometers was less than 2.42%, and the system was applied on neutron determination in the nuclear effect of acoustic cavitation (NEAC). And the neutron detection sensitivity formula of BF3 proportional counter was derived. The detection sensitivities of the BF3 detectors were calibrated by 2.45 MeV neutrons which were emitted from deuterium titanium target bombarded by deuterium ions of high-voltage accelerator. The sonofusion neutron signals were determined used the BF3 proportional counters, and the effective neutron emissivity estimations were of about 7.0×104 to 8.0×105 s-1.