LI Qi, WANG Shi-lian, ZHANG Xin-jun, FAN Yuan-qing, CHANG Yin-zhong, JIA Huai-mao, CHEN Zhan-ying, LIU Shu-jiang, ZHAO Yun-gang. Measurement of 7Be in CTBT Radionuclide Stations Verification Samples[J]. Atomic Energy Science and Technology, 2012, 46(12): 1498-1501. DOI: 10.7538/yzk.2012.46.12.1498
Citation: LI Qi, WANG Shi-lian, ZHANG Xin-jun, FAN Yuan-qing, CHANG Yin-zhong, JIA Huai-mao, CHEN Zhan-ying, LIU Shu-jiang, ZHAO Yun-gang. Measurement of 7Be in CTBT Radionuclide Stations Verification Samples[J]. Atomic Energy Science and Technology, 2012, 46(12): 1498-1501. DOI: 10.7538/yzk.2012.46.12.1498

Measurement of 7Be in CTBT Radionuclide Stations Verification Samples

  • The Comprehensive Nuclear-Test-Ban Treaty (CTBT) International Monitoring System (IMS) radionuclide network, including 80 stations, was designed to monitor nuclear test all over the word. The stations should be certified by CTBT to comply with the technical requirements. Among the certification tests carried out at each station, parallel sampling with portable particle sampler is performed. The verification samples from the portable air sampler will be analyzed at radionuclide laboratories to calibrate the equipments of stations. According to technical requirement of radionuclide station certification, this article deduced the formula of 7Be radioactivity concentration. The preparation and analysis procedure of verification samples were established. As part of certification of radionuclide stations RN42 and RN56, the verification samples from these two stations were measured by HPGe γ spectrometer in Beijing Radionuclide Laboratory.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return