TAN Yan-liang, WEI Nan-nan, SHAN Jian, XIAO De-tao, YUAN Hong-zhi. Method for Determination of 220Rn Concentration Reference Value by Low-pressure Scintillation Cell[J]. Atomic Energy Science and Technology, 2013, 47(7): 1277-1280. DOI: 10.7538/yzk.2013.47.07.1277
Citation: TAN Yan-liang, WEI Nan-nan, SHAN Jian, XIAO De-tao, YUAN Hong-zhi. Method for Determination of 220Rn Concentration Reference Value by Low-pressure Scintillation Cell[J]. Atomic Energy Science and Technology, 2013, 47(7): 1277-1280. DOI: 10.7538/yzk.2013.47.07.1277

Method for Determination of 220Rn Concentration Reference Value by Low-pressure Scintillation Cell

  • The accurate measurement of 220Rn is important for dosimetry evaluation of natural radiation exposure. The detection efficiencies of normal scintillation cell to different energy α particles decayed from 222Rn, 220Rn and their daughters are different for the large volume of normal scintillation cell. It was proposed that the range of α particles with minimum energy decayed from 222Rn could be longer than the longest distance between any two points in the scintillation cell by reducing the air pressure of scintillation cell. Thus, the detection efficiencies of the low-pressure scintillation cell to α particles decayed from 222Rn, 220Rn and their daughters are same. The detection efficiency of low-pressure scintillation cell to α particles can be obtained by standard radon chamber, and then the 220Rn concentration reference value can be determined by low-pressure scintillation cell.