Performance of LaBr3∶Ce/LaCl3∶Ce Scintillation Detector in Current Mode
-
Graphical Abstract
-
Abstract
LaBr3∶Ce and LaCl3∶Ce which exhibit superior ability in detection of γ-rays have been studied extensively in recent years. The properties for both of the LaBr3∶Ce scintillation detector and the LaCl3∶Ce scintillation detector were studied. The relative sensitivities to incident γ-rays of different energy were obtained by calculation. The sensitivities of detectors to the 137Cs and 60Co sources were measured and the sensitivity ratio of γ-ray to neutron was inspected. The result shows that the LaBr3∶Ce and LaCl3∶Ce scintillation detectors with high sensitivities to γ-rays and good γ/n discrimination are very appropriate for measurement in the mixed field by neutrons and γ-rays.
-
-