XU Guang-duo, WANG Li-qiang, TONG Jian-min. Research on Calibration Model of X-ray Thickness Gauge[J]. Atomic Energy Science and Technology, 2014, 48(5): 925-929. DOI: 10.7538/yzk.2014.48.05.0925
Citation: XU Guang-duo, WANG Li-qiang, TONG Jian-min. Research on Calibration Model of X-ray Thickness Gauge[J]. Atomic Energy Science and Technology, 2014, 48(5): 925-929. DOI: 10.7538/yzk.2014.48.05.0925

Research on Calibration Model of X-ray Thickness Gauge

  • A new model which describes the attenuation of X-ray in the X-ray thickness gauge was proposed. Based on measured data of several calibration plates whose thickness is already known, this model was compared with other gauging models mentioned in other papers according to different numbers of data points, different data point distributions and other situations. The calculated values with the new model are in good agreement with experimental data. When the number of data points decreases and the distribution of data points changes, the new model maintains a hign accuracy and presents a better numerical stability than other models.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return