LV Xiao-xia, CHEN Xi-lin, YAO Shun-he. Development of TDCR Liquid Scintillation System[J]. Atomic Energy Science and Technology, 2014, 48(11): 2133-2136. DOI: 10.7538/yzk.2014.48.11.2133
Citation: LV Xiao-xia, CHEN Xi-lin, YAO Shun-he. Development of TDCR Liquid Scintillation System[J]. Atomic Energy Science and Technology, 2014, 48(11): 2133-2136. DOI: 10.7538/yzk.2014.48.11.2133

Development of TDCR Liquid Scintillation System

  • The liquid scintillation counting system recently constructed for absolute activity measurement was introduced, and it uses TDCR method to measure activity of β nuclide. The system is equipped with three PMTs connected to a de-focusing voltage divider base developed by ourselves. After varying the efficiency by de-focusing of the PM tube, the activity of the 3H source was determined. The analysis of uncertainty and discussion of measurement result were given. The activity determined with TDCR is in agreement with certified value within the range of 0.9%. The combined uncertainty of results is 0.6% (k=1).
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return