Calculation and Experiment of Detective Efficiency of LaBr3:Ce Scintillation Detector
-
Graphical Abstract
-
Abstract
The absorption efficiency and peak-to-total ratio of LaBr3:Ce scintillators in different sizes were calculated by the Monte-Carlo method. The detection efficiency of a 50 mm×10 mm LaBr3:Ce detector was calibrated using three point sources 137Cs, 60Co and 152Eu. The calculation result shows a good agreement with the experiment result. The efficiency curve of the LaBr3:Ce detector to γ rays from 0 to 10 MeV was given by combining the calculation results with the experiment results. In this way, the detect efficiency of LaBr3:Ce detector could be acquired even there are not enough γ sources for calibration.
-
-