GAO Fei, XIAO Xue-fu, NI Ning. On-site Calibration Technology for Fixed Environmental Gamma Radiation Ratemeter[J]. Atomic Energy Science and Technology, 2015, 49(2): 212-217. DOI: 10.7538/yzk.2015.49.02.0212
Citation: GAO Fei, XIAO Xue-fu, NI Ning. On-site Calibration Technology for Fixed Environmental Gamma Radiation Ratemeter[J]. Atomic Energy Science and Technology, 2015, 49(2): 212-217. DOI: 10.7538/yzk.2015.49.02.0212

On-site Calibration Technology for Fixed Environmental Gamma Radiation Ratemeter

  • The fixed environmental gamma radiation ratemeter is the main equipment to undertake environmental continuous monitoring tasks and not easy to move for metrology laboratory calibration, and the submission of a longer period affects the continuity of the continuous monitoring data. For the on time calibration of fixed environmental gamma radiation ratemeter, combined with the Monte Carlo method, the energy-compensated high-pressure ionization chamber and portable 137Cs irradiation facility were developed. Using natural background radiation (including cosmic rays and terrestrial gamma rays) and 137Cs gamma rays reference radiation field, the on-site calibration for fixed environmental-gamma radiation ratemeter was carried out. The results show that the environment intercomparison and onsite irradiation can solve the calibration problem of fixed environmental gamma radiation ratemeters, and the relative deviation of calibration factors obtained from on-site calibration and in standard laboratory is less than 5%.