Analysis of Trace Hydrogen and Deuterium in Helium by High-precision Gas Chromatograph
-
Graphical Abstract
-
Abstract
It’s necessary to establish a high-precision analytical method of gas chromatography for achieving the rapid detection of trace components in fuel cycle system of fusion reactors. H2 and D2 in high purity helium were separated with a handmade cryogenic capillary column of Al2O3, and measured by discharge ionization detector. The results show that the detect limit of this method is less than 1×10-8, retention time is less than 180 s, the relative standard deviation of area response of H2 and D2 is not more than 1.0%, and the resolution of H2 and D2 is above 1.0. This method is sensitive and fast, and can be applied to fuelling system and isotope separation system of fusion reactors for tritium safety analysis and accurate measurement.
-
-