Citation: WANG Zilu, LI Yuqi, ZHANG Ming, TUO Fei, YANG Zhijie, LI Zeshu, LIANG Juncheng. Radioactivity Concentration Measurement of 85Kr Using Length-compensated Internal Gas Proportional Counting Method[J]. Atomic Energy Science and Technology, 2019, 53(12): 2465-2470. DOI: 10.7538/yzk.2019.youxian.0259

Radioactivity Concentration Measurement of 85Kr Using Length-compensated Internal Gas Proportional Counting Method

  • The radioactivity concentration absolute measurement of radioactive noble gas is the basis for the transmission of its quantity value. In order to determinate the value of 85Kr radioactivity concentration, the absolute measurement of 85Kr radioactivity concentration was carried out based on the length-compensated internal gas proportional counting method. The results show that the radioactivity concentration of 85Kr is 10.64 Bq/mL at pressure of 56.15 kPa and working voltage of 1 675 V and 1 700 V, and the combined standard uncertainty is 0.60%. The combined standard uncertainty is equivalent to the result reported by the international advanced nuclide metrology laboratory.