Abstract
In this paper, a simultaneous and rapid analysis method of 89Sr and 90Sr was established by proportional counter to measure twice after separation using Sr specific resin. The effects of 90Y growth factor and 89Sr/90Sr activity ratio on the measured values of 89Sr and 90Sr were studied. The results show that increasing 90Y growth factor is more conducive to obtain the activities of 89Sr and 90Sr. This method was applied to the analysis of 89Sr and 90Sr in samples with 89Sr/90Sr activity ratio ranging from 0.22 to 21.8, and the relative deviation between the measured value and the standard value is less than ±30%. The uncertainty analysis shows that the uncertainty of one nuclide will increase with activity ratio or activity of another nuclide. The maximum relative deviation of 89Sr and 90Sr between the measured value and the standard value is 7.44%, which proves the accuracy of this method by the analysis of IAEA proficiency test samples.
-
-