MA Lina, SONG Lijuan, YANG Yonggang, WANG Yadong, MA Yan, DAI Xiongxin. Method for Determination of 55Fe in Urea 14C Capsule[J]. Atomic Energy Science and Technology, 2021, 55(7): 1253-1259. DOI: 10.7538/yzk.2020.youxian.0530
Citation: MA Lina, SONG Lijuan, YANG Yonggang, WANG Yadong, MA Yan, DAI Xiongxin. Method for Determination of 55Fe in Urea 14C Capsule[J]. Atomic Energy Science and Technology, 2021, 55(7): 1253-1259. DOI: 10.7538/yzk.2020.youxian.0530

Method for Determination of 55Fe in Urea 14C Capsule

  • 55Fe in urea 14C capsule should be analyzed, so that the controllability and safety of pharmaceutical can be guaranteed. An analysis method of 55Fe was established for the urea 14C capsule. The method is that the samples are first purified twice by AGMP-1 anion exchange resin, and then 55Fe is determined by liquid scintillation counter (LSC) and the limit value of impurity radionuclide 55Fe in urea 14C capsule is calculated, and the specificity, durability and detection limit of the method are verified. The results show that activity limit of the impurity radionuclide 55Fe is less than or equal to 0.1% of the activity of 14C, the decontamination factor for 55Fe with 14C is higher than 105, the chemical recovery of Fe is 57%-67%, and the detection limit for 1 h is 0.4 Bq.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return