LI Qi, WANG Shi-lian, XIE Feng, HE Xiao-bing, FAN Yuan-qing, SHI Quan-lin, CHANG Yong-fu. Measurement of Emission Probability of 87Kr[J]. Atomic Energy Science and Technology, 2006, 40(增刊): 113-116.
Citation: LI Qi, WANG Shi-lian, XIE Feng, HE Xiao-bing, FAN Yuan-qing, SHI Quan-lin, CHANG Yong-fu. Measurement of Emission Probability of 87Kr[J]. Atomic Energy Science and Technology, 2006, 40(增刊): 113-116.

Measurement of Emission Probability of 87Kr

  • The relative gamma ray emission probabilities of 87Kr were measured by HPGe detector. The activity concentration of 87Kr was determined by using internal gas proportional counting system. The 402.6 keV γ-ray emission probability (49.9±1.2)% was obtained experimentally.