~(99)Mo-~(99m)Tc平衡源中~(99m)Tc和~(99)Mo强度比测定

MEASUREMENT OF THE RATIO OF ~(99m)Tc AND ~(99)Mo INTENSITY IN ~(99)Mo-~(99m)TC BALANCED SOURCE

  • 摘要: 本文用4πβ-γ符合方法中的外推吸收法测定了~(99)Mo-~(99m)Tc平衡源中的总“β”强度,用准确标定过的~(99m)Tc标准源刻度好效率的S-80多道高纯锗γ谱仪测定平衡源中~(99m)Tc的绝对强度,从而测定了~(99m)Tc和~(99)Mo的强度比。结果为0.974±0.0023。并将结果和编评值比较,为~(99)Mo的绝对标定提供了~(99m)Tc的修正量为0.1088±0.0023的实测值。

     

    Abstract: The measurement of the total β intensity of a ~(99)Mo-~(99m)Tc balanced source byextrapolation absorption method of 4πβ-γ coincidence and the measurement ofabsolute intensity of ~(99m)Tc in the balanced source by a S--80 MCA HPGE γspectrometer with the efficiency calibrated by an accurately standardized sourceof ~(99m)Tc are described. The ~(99m)Tc/~(99)Mo intensity ratio is found to be 0.974±0.023. This result is compared with the compilation value. Thus a experimentallymeasured correction factor of ~(99m)Tc is 0.1088±0.0023 for the absolute standar--dization of ~(99)Mo.

     

/