Abstract:
A 4πβ γ anti coincidence system is developed for absolute radioactivity measurement.In this system,a dual ADC live time HPGe γ spectroscopy is used as the γ ray detector,and a 4πβ proportional counter is employed for the β detector. The activities of 60 Co and 134 Cs are measured by this system. The combined standard uncertainty of 60 Co and 134 Cs is 0.24% and 0.33%,respectively. The results are compared with that from 4πβ γ coincidence counting system of NIM and CIAE and the agreement of these results is satisfactory.