γ射线探测器能量响应标定技术
Technique of Calibrating Energy Response of γ-ray Detector
-
摘要: 叙述了γ射线探测器能量响应的标定原理和方法,即利用Compton散射将强60Co源的1.25 MeVγ射线转换为0.36~1.02 MeV(25°~90°)范围内任意能量的系列单能γ射线;采用辐射屏蔽技术和提高能量分辨率方法对γ射线探测器进行了该能区的能量响应标定;通过MCNP程序对探测器的能量沉积趋势和不同散射角散射γ射线的能量分布进行模拟计算。结果表明:该标定系统信噪比达到了20∶1左右,能量分辨率约为6%,0.66 MeV的灵敏度与标准137Cs源直照标定的灵敏度基本一致。Abstract: The paper introduces a method and principle for calibrating energy response of γ-ray detector. The energy response may be obtained by means of transforming γ-ray from 60Co into a series of mono-energetic γ-ray in the range of 0.36~1.02 MeV, shielding radiation-interference and improving energy resolution. The energy distributions of scattered γ-ray at different angles and corresponding energy deposited in detector were simulated by MCNP code. The results show that the signal-to-noise of the system is about 20∶1, the energy resolution is about 6%,and the sensitivity of detector to 0.66 MeV γ-ray is in agreement with the one to the γ-ray from 137Cs.