用反符合方法测量131I和133Ba的比活度

Specific Activity Measurement of 131I and 133Ba by Anti-coincidence Method

  • 摘要: 131I是核医学应用中较为重要的核素之一;133Ba是1种电子俘获核素,对HPGe γ谱仪能量刻度和效率校准具有重要意义。利用4πβ(PPC)-γ(HPGe)反符合测量装置对上述两种核素的活度进行了反符合测量方法研究,所得到的比活度测量结果分别为1 794.8(1±0.6%)kBq•g-1k=1)和502.8•(1±0.6%)kBq•g-1k=1),与归一到相同参考时刻4πβ(PC)-γ(NaI(Tl))符合效率外推法的测量结果1 797.6(1±0.5%)kBq•g-1k=1)和504.9(1±0.6%)kBq•g-1k=1)均在不确定度范围内一致。

     

    Abstract: 131I is a very important nuclide applied in nuclear medicine. 133Ba, an EC decay nuclide, plays a significant role in the efficiency calibration of HPGe detector. The activity measurement method for the above two kinds of nuclides was studied by the 4πβ(PPC)-γ(HPGe) anti-coincidence counting. The specific activity measurement result for 131I sample is 1 794.8(1±0.6%)kBq•g-1k=1)and for 133Ba is 502.8(1±0.6%)kBq•g-1(k=1). Compared with the results measured by 4πβ(PC)-γ(NaI(Tl)) coincidence efficiency extrapolation methods at the same reference time, 1 797.6(1±0.5%)kBq•g-1k=1)and 504.9(1±0.6%)kBq•g-1k=1), it is in good agreement within the uncertainties.

     

/