自制个人中子气泡探测器特性的批内一致性
In-Batch Homogeneity of Personal Home-Made Neutron Bubble Detectors
-
摘要: 采用252Cf中子源和2×1.7 MV串列加速器产生的14.8 MeV单能中子源,对自制的12批中子气泡探测器的批内一致性进行测试。结果表明,在室温条件下,同批次制作的气泡探测器性能符合ISO标准。Abstract: The in-batch homogeneity of personal home-made neutron dosimeter bubble detector was studied by using 252Cf neutron source and 14.8 MeV mono-energy neutron with ISO guidelines as criteria. The result shows that the in-batch homogeneity of personal home-made neutron dosimeter bubble detector is qualified for monitoring doses from both 252Cf and 14.8 MeV neutron.