Abstract:
In order to achieve the high efficiency and uniform sensitivity for neutrons with widely different energies, the efficiency of long boron trifluoride proportional counter imbedded in polyethylene moderator was simulated by MCNP code. The result shows that detective efficiency would increase with increasing moderator radius and response curve at higher energy would be ameliorated through adjusting the thickness of front moderator. Also we calculated the relative efficiencies for different energy of a detector whose efficiencies were calibrated on an accelerator. The simulated efficiency for D-D neutrons (2.4 MeV) is 75% of the efficiency for D-T neutrons (14.1 MeV), which is approximately agreed with experimental data, 61%. The validity of the simulated model was proved by the consistent results between calculation and experiment data.