An Efficient Fast Recovery Scheme for Nuclear Quadrupole Resonance Single Coil Probe

  • 摘要: 介绍了一种用于主动减小核四极共振(NQR)探头恢复时间的Q值变换电路,它利用MOSFET控制变Q电路与探头电路的耦合,并通过变压器增加与电感线圈串联的阻抗值以达到改变Q值的目的。此电路不产生新的振铃,对接收NQR的分辨率也不造成损失。实验结果表明,电路能有效缩短探头的恢复时间,同时还可提高信噪比。

     

    Abstract: An active Q-switching circuit using a transformer for reducing probe recovery time just after the RF excitation was described. MOSFET controls the coupling and uncoupling to the Q-switching circuit from the probe. It doesn’t create new ringing and doesn’t operate during the reception of nuclear quadrupole resonance signal and there’s no loss in sensitivity. Experimental results prove that the probe recovery time is reduced and SNR is increased by this circuit.

     

  • [1] 孙汉城,李广将.检查隐藏违禁品的核技术[J]. 同位素,2005,18(1-2):110-116.
    SUN Hancheng, LI Guangjiang. Nuclear tech-niques for detecting hidden contraband[J]. Journal of Isotopes, 2005, 18(1-2): 110-116(in Chinese).
    [2] 石光明,徐更光,王廷增,等.核四极共振在炸药探测上的应用[J]. 火炸药学报, 2004, 27(3):70-73.
    SHI Guangming, XU Gengguang, WANG Tingzeng, et al. Applications of nuclear quadrupole resonance in explosives detection[J]. Chinese Journal of Explosives & Propellants, 2004, 27(3): 70-73(in Chinese).
    [3] ROW E M D, SMITH J A S. The detection of abandoned landmine[C] ∥Proceedings of EUREL International Conference. London: IEE Conference Publication, 1996: 62-66.
    [4] RUDAKOV T N, MIKHALTSEVICH V T, SELCHIKHIN O P. The use of multi-pulse nuclear quadrupole resonance techniques for the detection of explosives containing RDX[J]. J Phys D: Appl Phys, 1997, 30: 1 377-1 382.
    [5] RUDAKOV T N, HAYES P A, MIKHALTSEVITCH V T. “Magic echo” multi-pulse sequence in nitrogen-14 NQR[J]. Physics Letters A, 2004, 330: 280-285.
    [6] CLARK W G. Pulsed nuclear resonance apparatus[J]. Rev Sci Instrum, 1964, 35(3): 316-333.
    [7] HOULT D I. Fast recovery, high sensitivity NMR probe and preamplifier for low frequencies[J]. Rev Sci Instrum, 1979, 50: 193-200.
    [8] MILLER J B, SUITS B H, GARROWAY A N. Interplay among recovery time, signal, and noise: Series- and parallel-tuned circuits are not always the same[J]. Concepts Magn Reson, 2000, 12: 125-136.
    [9] SULLIVAN N S, DESCHAMPS P, NEEL P, et al. Efficient fast-recovery scheme for NMR pulse spectrometers[J]. Rev Phys Appl, 1983, 18: 253-261.
    [10] 王兆安,张明勋.电子电力设备设计和应用手册[M]. 北京:机械工业出版社,2002.
计量
  • 文章访问数:  1279
  • HTML全文浏览量:  0
  • PDF下载量:  1198
  • 被引次数: 0
出版历程
  • 收稿日期:  2007-12-09
  • 修回日期:  2008-03-20
  • 刊出日期:  2009-05-19

目录

    /

    返回文章
    返回