多重性探测器中子衰减时间的标定
Neutron Decay Time Calibration of Multiplicity Counter
-
摘要: 采用Rossi-α方法直接测量了多重性探测器中子衰减时间,并采用几个不同强度的锎源和不同质量的钚部件进行了Rossi-α验证测量,准确获得了某个多重性探测器的中子衰减时间。Rossi-α直接测量相比改变符合时间宽度间接标定的传统方法更直接可靠,且可在钚部件无损测量中作为辅助监测手段。Abstract: Rossi-α method was proposed to directly measure the neutron decay time of the multiplicity counter. Based on this method multiplicity counter’s neutron decay time was measured with several different density 252Cf and different mass Pu metal.