Abstract:
The calculation of relevant parameters of the button beam position monitor (BPM) at Hefei Light Source (HLS) was presented. The horizontal and vertical sensitivity were calculated. The intrinsic resolution and transmission power were calculated under certain parameters. The high precision beam position was simulated with the optimized detector installation location parameters. The measured data were processed by the difference/sum and log-ratio method to obtain the intuitive mapping figure and fitting polynomial. The sum signal was simulated and fitted. The variety range of normalized sum signal was less then 5×10—3 when the beam displacement changed from (0 mm, 0 mm) to (4 mm, 4 mm).