Abstract:
Contrast sensitivity describes the ability of the micro-CT system to reproduce object composition, which mainly depends on the noise of the image. The method of measuring the contrast sensitivity was introduced. A phantom with 50 mm diameter was measured on the system 225 kV-3D-micro ICT and it is found the contrast sensitivity is about 1% at the range of 0.3 mm×0.3 mm×0.3 mm. In addition, the system stability, beam hardening and the reconstruction parameter which affect the measurement result were studied.