闪烁薄膜探测器X/γ射线能量响应研究

Investigation About Energy Response of Scintillation Film Detector to X/γ Rays

  • 摘要: 为准确评估闪烁薄膜探测器(OSFD)在脉冲裂变中子参数测量中γ射线对测量结果的影响,利用Geant4程序对闪烁薄膜探测器的X(或γ)射线响应进行模拟,结合半经验电子发光效率曲线,获得探测器能量响应理论曲线,利用662 keV和1.25 MeV的单能γ射线源,以及窄谱剂量标准的48~208 keV准单能X射线能点对探测器响应进行刻度。本研究结果为闪烁薄膜探测器结构的改进,以及探测器在宽能谱脉冲X射线场测量中的应用提供了重要依据。

     

    Abstract: To obtain the impact of γ rays for the organic scintillation film detector (OSFD) in the field of pulsed fission neutron detection, the response of the OSFD to X or γ rays was studied with the Geant4 program. Combining with a semi-empirical scintillation efficiency function, the energy response function of the OSFD to γ rays was obtained. To verify the simulation result, the sensitivities of an OSFD employing 0.3 mm film to some discrete energy points were tested experimentally with 137Cs(662 keV) and 60Co (1.25 MeV) γ ray sources, as well as some discrete semi-monoenergetic X ray points (48 to 208 keV) of a standard narrow band X ray device. The research is instructive for the improvement of the detector structure, and the results are valuable for the application of the detector in measuring pulsed X rays with wide energy distribution.

     

/