Abstract:
The determination of carbon content in LiH was studied with HCS-878F high-frequency infrared C-S measurer. Various experiment conditions such as heating temperature, type and dosage of flux and sampling amount were studied. The results show that hydrogen in the example can be released during burning by increasing the power of the instrument step by step without fulmination. To avoid the influence of hydrogen released on the result of determination, a device for translating CO into CO