Abstract:
The accurate measurement of
220Rn is important for dosimetry evaluation of natural radiation exposure. The detection efficiencies of normal scintillation cell to different energy α particles decayed from
222Rn,
220Rn and their daughters are different for the large volume of normal scintillation cell. It was proposed that the range of α particles with minimum energy decayed from
222Rn could be longer than the longest distance between any two points in the scintillation cell by reducing the air pressure of scintillation cell. Thus, the detection efficiencies of the low-pressure scintillation cell to α particles decayed from
222Rn,
220Rn and their daughters are same. The detection efficiency of low-pressure scintillation cell to α particles can be obtained by standard radon chamber, and then the
220Rn concentration reference value can be determined by low-pressure scintillation cell.