基于反应堆中子照相孔道参数初步实验研究

Preliminary Experimental Study on Parameters of Conduct for Neutron Photography Based on Reactors

  • 摘要: 在医院中子照射器反应堆现有的水平热中子孔道的基础上,设计了一个适用于中子照相的热中子准直器。准直器入口2 cm或4 cm,出口直径为10 cm,长度为115 cm,最大准直比为58,偏移角为4.1°。对设计的热中子孔道进行了初步实验研究,结果显示,满功率运行时准直器出口处热中子注量率为1.0×106 cm-2•s-1,γ本底剂量率平均为11.65 mSv/h,达到了热中子照相的要求。

     

    Abstract: Based on the available thermal neutron beam of In-hospital Neutron Irradiator (IHNI), a collimator was designed and constructed for the purpose of neutron radiography. The diameter of the collimator entrance was 4 or 2 cm while the exit was 10 cm. The most collimation ratio was 58 and the divergence angle was 4.1°. The parameters of the neutron beam were studied by experiments. At full power of IHNI, the neutron fluence rate at the exit of the collimator is 1.0×106cm-2•s-1, while the average γ dose rate is 11.65 mSv/h, which could meet the requirement of neutron radiography.

     

/