Abstract:
To obtain the n/γ discrimination ability of different digital pulse shape discrimination methods, four methods (rising time method, charge comparison method, pulse gradient analysis and frequency gradient analysis) in americium-beryllium mixed radiation fields were demonstrated. The signals from EJ-301 and BC501A scintillator detectors were digitized using oscilloscope. A comparison was taken among the four discrimination methods. The discrimination results of the four methods in liquid scintillator detectors show that the rising time method is the best and it provides a good choice in real-time n/γ discrimination system.