L吸收边法测定台架实验中的铀

Determination of Uranium in Bench Test by L-absorption Edge Method

  • 摘要: 采用自制的L吸收边分析装置,建立了测量20~200 g/L铀浓度的分析方法。通过实验确定了测量中铀的L吸收边左拟合区间为1 659~1 856道,右拟合区间为2063~2280道,铀的L吸收边为1995道。9 mol/L以下的硝酸、10 g/L以下的Al以及10 g/L以下的Fe对铀测量无显著影响。在铀的工作曲线范围内,其测量精密度均优于0.5%,稳定性好。利用本装置测量铀回收处理台架实验中的部分样品,取得了满意结果。

     

    Abstract: By independent research L-absorption edge densimeter, an analytical method for uranium sample with the concentration from 20 g/L to 200 g/L was developed. The fitting area for uranium measurement was determined through experiment. The left fitting area was 1659-1856 channel, and the right one was 2063-2280 channel. The uranium L-absorption edge was at 1 995 channel. The results show that the influence of HNO3 concentration lower than 9 mol/L, Al and Fe concentration lower than 10 g/L was negligibly small. The uranium measurement precision is better than 0.5%, and the instrument stability is good. Some samples in bench test of uranium recovery were determined. The results are satisfactory.

     

/