Abstract:
By independent research L
Ⅲ-absorption edge densimeter, an analytical method for uranium sample with the concentration from 20 g/L to 200 g/L was developed. The fitting area for uranium measurement was determined through experiment. The left fitting area was 1659-1856 channel, and the right one was 2063-2280 channel. The uranium L
Ⅲ-absorption edge was at 1 995 channel. The results show that the influence of HNO
3 concentration lower than 9 mol/L, Al and Fe concentration lower than 10 g/L was negligibly small. The uranium measurement precision is better than 0.5%, and the instrument stability is good. Some samples in bench test of uranium recovery were determined. The results are satisfactory.