Abstract:
In order to meet with the requirement of electromagnetic diagnosis to the J-TEXT Tokamak, a remote control integrator system was developed. With modular design method, the integrator system is composed of the integrator cards, a control card, a linear power card and the BNC interface cards, and it uses the PC control software to conduct network control. An integrator system provides 32 integrator channels, and all integral channels have four kinds of integral time constants for remote selection and provide three kinds of integrator running control methods. According to laboratory and J-TEXT field testing, it shows that the output voltage range is -10-10 V, output noise is not more than 5 mV, and for the four kinds of integral time constants, the integral output drifts are all less than 5 mV within 100 s for each integrator channel.