Abstract:
A new method of continuous measurement of the concentration of
220Rn progeny without replacing membrane frequently was established based on α spectroscopy. A comparison study between the new method and the standard method for determination
220Rn progeny reference levels was carried out in the radon laboratory of the University of South China. The results show that under the high level measurement environment of
220Rn progeny, the relative difference of the two method is 5.0% for ThB and 7.0% for ThC during eight measurement periods. Under the low level measurement environment of
220Rn progeny, ThB relative difference of the two methods is controlled within 4.5% and that of ThC is controlled at about 10.0% during four measurement periods. The total uncertainty of ThB and ThC is controlled at about 5% for measurement
220Rn progeny’s level of 100 Bq/m
3 in the environment, respectively. The effect of aerosols deposited on the membrane surface on membrane self-absorption can be ignored in the process of sampling. The new method provides accuracy, quick and continuous measurement for
222Rn/
220Rn progeny.