Abstract:
The mass resolution of Beijing Radioactive Ion Beam Facility Isotope Separator On-line (BRISOL) is 20 000 for the separation of isobar beam. The field measurements and field optimizing were presented in this paper in detail. To reach the high mass resolution power of 20 000, a bi-energy system with two small dipoles and two big dipoles was used to reduce energy dispersion which is too large to get high mass resol ution power. The magnetic field uniformity of the analysis magnet must be better than 5×10
-5. The surface coils were set on pole face to improve the magnetic field homogeneity. The integral field uniformity with surface coil was optimized from 3×10