TEPC中子和γ分辨技术研究

Study of Neutron and γ Discrimination with TEPC

  • 摘要: 混合辐射场中子剂量、剂量当量的测量需进行中子、γ分辨。依据各种辐射沉积线能的不同,组织等效正比计数器(TEPC)具有一定的中子、γ分辨能力。本文采用自制的圆柱形TEPC在5SDH-2加速器单能中子辐射场进行了微剂量谱测量,对其中子、γ分辨技术进行了探讨分析。采用137Cs纯γ辐射微剂量谱匹配法,在252Cf、241Am-Be中子辐射场进行了中子、γ分辨研究。分辨后的中子剂量当量与约定真值一致性较好,表明TEPC用于中子、γ混合辐射场的吸收剂量、剂量当量测量是可行的。

     

    Abstract: Measurements of neutron dose and dose equivalent in mixed radiation field should discriminate neutron and gamma. According to the characteristic of neutron and gamma linear energy induced in detector, the tissue equivalent proportional counter (TEPC) could discriminate γ fraction from total spectrum measured in mixed radiation field. In this paper, the microdosimetric spectra were measured with homemade cylindrical TEPC in monoenergy neutron radiation field produced by the 5SDH-2 accelerator. Based on the pure γ microdosimetric spectrum measured in the 137Cs γ reference radiation field, the neutron and γ discriminations were studied in 252Cf, 241Am-Be neutron reference radiation fields. The pure neutron dose equivalent rates of 252Cf, 241Am-Be sources agree well with reference ambient dose equivalent. The results show that it is practicable to measure absorbed dose and dose equivalent with TEPC in neutron and γ mixed field.

     

/