中国实验快堆固定式γ探测器现场校准技术研究
Research on On-site Calibration Technology of Fixed γ Detector at China Experimental Fast Reactor
-
摘要: 通过研究γ探测器实验室校准的技术条件,在考虑中国实验快堆(CEFR)现场校准的各种影响因素后,根据CEFR实际情况组装一套校准专用装置,并选择替代法为现场校准方法,选择合适的参考仪器对CEFR厂房内固定式γ探测器进行现场校准实验。结果表明,采用CEFR校准专用装置和建立的现场校准方法可满足固定式γ探测器的现场校准要求。Abstract: Based on numerous studies of γ detector calibration in laboratory and actual situation at China Experimental Fast Reactor (CEFR), a set of special calibration device which is suitable to calibrate fixed γ detector at CEFR was designed and manufactured, and the substitution method was used for the calibration of CEFR fixed γ detector. The results show that the special calibration device and the calibration technology satisfy the requirement of CEFR fixed γ detector on-site calibration.