准阻抗匹配传输式的外移动式束流探测器标定装置的研发与使用

Development and Application of External Moving Type Calibration Device for Beam Detector Based on Quasi-impedance Matching Transmission

  • 摘要: 为了标定并测量Rhodotron电子加速器中束流的横向偏置,提出准阻抗匹配传输式的外移动式束流探测器标定装置,整个标定装置的外导体内径与内导体外径保持2.303的比例,以使电磁波在该标定装置传输中保持50 Ω的阻抗匹配。本文研究了标定装置的内导体外径对束流探测器频谱响应的影响,结果表明:改变内导体外径并不改变束流探测器的频率谐振峰位置。首次采用固定内导体并移动标定装置外部的束流探测器(外导体)以使标定装置保持准阻抗匹配,并对比了外移动式束流探测器标定装置与固定外导体而移动内导体标定装置对束流标定的影响,结果表明:无论束流偏置与否,两种标定装置的束流探测器频谱响应曲线高度重合,且外移动式束流探测器标定装置的阻抗匹配结果更优。最终,将外移动式束流探测器标定装置应用在实际束流标定上,证明该标定装置能用于Rhodotron电子加速器的束流标定。

     

    Abstract: The beam calibration results are affected by the impedance of the calibration device. If impedance matching is not maintained, reflected signals would be generated, which would lead to the superposition of reflected signals and excitation signals, therefore, the signals that stimulated BPM are not similar to the real beam signals. In addition, the unsealed outer conductor of some beam calibration devices may lead to leakage of excitation signals or detection of other electromagnetic signal noises in the laboratory, thus increasing the calibration and detection errors. In order to calibrate and measure the transverse bias of the beams in the Rhodotron electron accelerator more accurately, an external moving type beam calibration device based on quasiimpedance matching transmission was proposed, and its outer conductor was completely enclosed. The connection of the main components for the outer conductor is as follows: excitation signal input portfirst fixed straight line sectionfirst bellows sectionfirst impedance matching section-BPM-second impedance matching sectionsecond bellows section-second fixed straight line section-excitation signal output port. In order to study the calibration device, CST MicroWave Studio was used to establish the calibration device, and the beam bias was simulated by fixing the inner conductor rod of the calibration device and moving the beam detector (outer conductor) outside the calibration device, so it is called external moving type calibration device. The ratio between the inner diameter of the outer conductor and the outer diameter of the inner conductor is kept at 2.303, so that the impedance matching of the electromagnetic wave in the transmission of the calibration device is maintained at 50 Ω. The influence of outer diameter of inner conductor on the spectral response of detector was studied. The results show that changing the size of the inner rod does not change the resonant peak position of the beam detector. The inner conductor rod is fixed and the outer conductor is moved for the first time to maintain the quasiimpedance matching state of the calibration device. In the impedance matching section, the aperture of the two ends of the pipe can be changed to adapt to BPMs of different apertures. To calibrate BPMs of other aperture, only the impedance matching section and the inner conductor rod need to be redesigned, thus increasing the utilization rate of other equipments of the calibration device. The effect of external moving type calibration device on beam calibration was compared with that of fixed outer conductor and moving inner conductor calibration device. The results show that the detector spectral response curves of the two devices coincide highly regardless of whether the beam is offset or not, and the outer conductor moving calibration device has a better impedance matching. Finally, the outer conductor moving calibration device was applied to the actual beam calibration, which proves that the calibration device could be used for the calibration of Rhodotron electron beam.

     

/

返回文章
返回