狭缝法测量X射线管焦点尺寸的研究

Research on Slit Method for Measuring Focal Spot Size of X-ray Tube

  • 摘要: 目前EN 12543标准中的狭缝法广泛应用于工业用X射线管焦点尺寸的测量和标注。然而所需的测试设备——狭缝相机,由于其复杂和精密的设计要求,导致生产加工难度大、成本高且不利于推广使用。本文采用双钨合金圆柱代替狭缝相机,基于圆柱法中距离和位置参数校准方法,改进了狭缝法中焦点尺寸测量的计算公式,研究了不同狭缝大小对X射线管焦点尺寸测量的影响和变化规律。实验结果表明,使用间距与焦点尺寸相当的双钨合金圆柱测量COMET公司的MXR-225HP/11工业X射线管,也可以获得较为精确的测量结果,同时实验发现当狭缝宽度较大时,可以直接用来计算焦点测量装置的几何放大倍数,并且计算结果与圆柱法中使用距离和位置参数校准方法所测量的结果相一致。

     

    Abstract: The focal spot size of X-ray tube is one of the key factors affecting the imaging resolution, and it is also an important index to characterize the working performance of the X-ray source. Typically, after the X-ray tube is manufactured, the focal spot size must be measured and marked. Nowadays, the most widely used measurement standards for X-ray tube focal spot size in the world are IEC 60336, EN 12543, and ASTM E1165. Between various standards, there are differences in the focal spot size measurement methods, scope of application, loading factors and imaging equipment. Among these, the slit method of EN 12543 standard is mostly utilized in industrial X-ray tubes. However, due to its complex and precise design requirements, the test equipment, i.e., the slit camera, is difficult to produce and process, and the cost is high, which is not conducive to promotion and application. This paper studied the influence and variation of the measurement of the focal spot size at different slit sizes by using double tungsten alloy cylinders instead of slit camera. Based on the calibration of distance and position parameters in the cylinder method, the measurement equipment was conveniently positioned, and the distance between the focal spot and the translation mechanism was measured simultaneously, which improves the position accuracy of the focal spot size measurement. Additionally, the formula for determining the focal spot size in the slit method was improved, so that the slit method can be applied to a wide range of slit widths. In particular, the geometric magnification of the system can be measured directly when the slit width is large, skipping the step of measuring the distance between the focal spot and the translation mechanism. The experiment scheme has low requirements on the centering conditions of the equipment and the distribution of the X-ray source. In the data processing process, dark field correction and bright field correction were used to eliminate the problem of inconsistency in the response characteristics of each detector pixel, and the Wiener Filtering was performed on the signal output by the detector to deal with Gaussian noise and uniformly distributed noise. The experiment results show that relatively accurate measurement values can be obtained by measuring MXR-225HP/11 with double tungsten alloy cylinders with the same spacing as the focal spot size. The measurement relative error is less than 10%. According to international standards, when the measurement results are within plus or minus 10% of the theoretical value, it can be used as a method to determine the focal spot size. Therefore, the improved method in this paper is reliable and does not require high testing equipment, allowing it to be widely used. In addition, the geometric magnification of the focus measurement device can be directly calculated when the slit width is larger than the focal spot size, and the calculation results are consistent with the results measured in the cylinder method.

     

/

返回文章
返回